Product introduction: BM2007 calcium ferrum element measuring apparatus

 Product introduction: BM2007 calcium ferrum element measuring apparatus is the quantitative analysis apparatus of lime stone (lime powder) in building material, coal, metallurgy, petroleum, mining, chemistry, power plant desulfuration as well as limestone in coal refuse clinker brick adobe. It is mainly used to measure calcium and ferrum content in cement slurry, limestone (lime powder) and related products for generating station desulfuration. We control clinker quality by measuring calcium and ferrum content in slurry. At the same time, we can also analyze intake quantity of composite materials in cement so as to promptly supply data for proportion control.

BM2007 calcium ferrum element measuring apparatus adopts physical analysis method to guarantee quick analysis speed, high accuracy, large analyzing range width, good  repeatability, less human error, low labor intensity of operator, free pollution and other features. Free from any chemical agents, it conforms to environmental protection and energy-saving requirement. At the same time, the apparatus conforms to requirement of government standard (GB/T176-2008) cement chemical analysis method and X-ray fluorescence analysis instrument standard for (JC/T1085-2008) cement. 

Main feature: 

(1) Free from radioactive source, it is within the management range exempted by environmental protection department; 

(2) Large liquid crystal display and Chinese menu are convenient for operation; 

(3) Automatic audio alarm, keyboard printer in panel and operation in man-machine interactive mode; 

(4) Physical analysis method is adopted to avoid sample contact or damage during analysis process. It is unnecessary to adopt chemical agent; 

(5) Highly integrated instrument to guarantee powerful adaptive capacity to environment and interference-free feature; 

(6) Instrumental analysis system can be used independently or used together with microcomputer material proportioning system by connecting through RS232 or RS485 standard serial port; 

(7) Two elements are analyzed at the same time quickly. We adopt "continuous measurement method of standard sample" so as to obtain the result with high precision in 30s; 

(8) Strong software function and great data storage capability; Software function can be added as required by users to guarantee high cost performance; 

(9) After measurement is completed, "pull-out measurement tray" warning function is prompted within 8s (new functions are unavailable from other manufacturers)

Key technical indexes: 

◎ Measurement range CaO, Fe2O3 analysis range can be adjusted and Selected by the calibration curve method

◎ Measurement range widthCaO%maxCaO%min7%Fe2O3%maxFe2O3%min5%

Such as cement slurryCaO39%46%Fe2O30.015%

◎ Inherent error CaO:±0.18%Fe2O3:±0.10%

◎ Analysis precision SCaO0.09%,SFe2O30.05%

◎ Measurement time 30×n(n is a natural number)The recommended value is 60 seconds

◎ Operating conditions Ambient temperature540Relative Humidity:≤85%(30)Power supply220V±20V50Hz

◎ Power consumption of complete machine 60W

◎ Dimension and weight 480mm×370mm×125mm12kg