BM2011 silicon aluminium element measuring apparatus

 BM2011 silicon aluminium element measuring apparatus

Abstract: BM2011 silicon aluminium element measuring apparatus as the quantitative analysis apparatus is developed as required by cement enterprises. It is mainly used to measure SiO2 and Al2O3 content in slurry. We can use the apparatus to monitor change condition of SiO2 and Al2O3 during production process in teal time. Moreover, to promptly and accurately obtain SiO2 and Al2O3 content is equally important to control of cement production quantity by CaO and Fe2O3. Slurry rate value can be controlled during production process only by prompt and accurate obtaining of these four elements so as to improve slurry quantity. BM2011 silicon aluminium element measuring apparatus is widely applicable to metallurgy, mining, building materials and other industries. 

Main feature: 

(1) Free from radioactive source, it is within the management range exempted by environmental protection department; 

(2) Large liquid crystal display and Chinese menu are convenient for operation; 

(3) Automatic audio alarm, keyboard printer in panel and operation in man-machine interactive mode; 

(4) Physical analysis method is adopted to avoid sample contact or damage during analysis process. It is unnecessary to adopt chemical agent; 

(5) Highly integrated instrument to guarantee powerful adaptive capacity to environment and interference-free feature; 

(6) Instrumental analysis system can be used independently or used together with microcomputer material proportioning system by connecting through RS232 or RS485 standard serial port; 

(7) Two elements are analyzed at the same time quickly. We adopt "continuous measurement method of standard sample" so as to obtain the result with high precision in 30s; 

(8) Strong software function and great data storage capability; Software function can be added as required by users to guarantee high cost performance; 

(9) After measurement is completed, "pull-out measurement tray" warning function is prompted within 8s (new functions are unavailable from other manufacturers)

Key technical indexes: 

◎ Measurement range SiO2, Al2O3 analysis range can be adjusted, selected by the calibration curve method

◎ Measurement range widthSiO2maxSiO2 min15%Al2O3maxAl2O3min7%

◎ Inherent error SSiO20.07%SAl2O30.07%

◎ Analysis precision SCaO0.09%,SFe2O30.05%

◎ Measurement time 30×n(n is a natural number)The recommended value is 60 seconds

Temperature Stability : In the range of 5 ~ +40 , absolute drift: | SiO2 | 0.05%, Al2O3 | 0.05%;

◎ Operating conditions Ambient temperature540Relative Humidity:≤85%(30)Power supply220V±20V50Hz

◎ Power consumption of complete machine 60W

◎ Dimension and weight 480mm×370mm×125mm12kg